Rated conditions for gauge pressure and absolute pressure (from the gauge pressure series)
Measuring cell with inert
liquid
With extension to Zone 0
Conditions of use for gauge pressure and absolute pressure with
flush-mounted diaphragm
Installation conditions
Ambient temperature
Note
Measuring cell with sili-
cone oil filling
Measuring cell with inert
liquid
(various pressure clas-
ses)
Measuring cell with Neo-
bee (FDA-compliant)
Display
Storage temperature
Climate class
Condensation
Degree of protection in
accordance with EN 60
529
Degree of protection in
accordance with NEMA
250
Electromagnetic compatibility
SITRANS P DS III/P410 with PROFIBUS PA
Operating Instructions, 02/2016, A5E00053276-08
1 bar/100 kPa/3.6 psi
4 bar/400 kPa/58 psi
16 bar/1.6 MPa/232 psi
63 bar/6.3 MPa/914 psi
160 bar/16 MPa/2321 psi
400 bar/40 MPa/5802 psi
700 bar/70 MPa/10152 psi
Observe the temperature class in hazardous areas.
-40 ... +85 °C (-40 ... +185 °F)
1 bar/100 kPa/3.6 psi
4 bar/400 kPa/58 psi
16 bar/1.6 MPa/232 psi
63 bar/6.3 MPa/914 psi
160 bar/16 MPa/2321 psi
400 bar/40 MPa/5802 psi
700 bar/70 MPa/10152 psi
-10 ... +85 °C (14 ... 185 °F)
-30 ... +85 °C (-22 ... +185 °F)
-50 ... +85 °C (-58 ... +185 °F)
(with Neobee: -20 ... + 85 °C (-4 ...
+185 °F))
(with high-temperature oil: -10 ... +
85 °C (14 ... 185 °F))
Permitted
IP66, IP68
NEMA 4X
-40 ... +100 °C (-40 ... +212 °F)
-40 ... +100 °C (-40 ... +212 °F)
-40 ... +100 °C (-40 ... +212 °F)
-40 ... +100 °C (-40 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +60 °C (-4 ... +140 °F)
-40 ... +100 °C (-40 ... +212 °F)
-40 ... +100 °C (-40 ... +212 °F)
-40 ... +100 °C (-40 ... +212 °F)
-40 ... +100 °C (-40 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
-20 ... +100 °C (-4 ... +212 °F)
Technical data