Absolute pressure input (from the gauge pressure series)
for process temperature -20°C < ϑ
≤ 60°C (-4°F < ϑ ≤ +140°F)
for process temperature 60°C < ϑ
≤ 100°C (max. 85°C for measuring
cell 30 bar) (140°F < ϑ ≤ 212°F
(max. 185°F for measuring cell
435 psi))
Upper measuring limit
Start of scale value
Order the nominal measuring range with the order option Y01 for PROFIBUS PA or FOUNDATION Fieldbus.
1)
Absolute pressure input (from the differential pressure series)
Measured variable
Span (continuously adjustable) or measur-
ing range and max. operating pressure (in
accordance with 97/23/EC Pressure
Equipment Directive)
Lower measuring limit
Measuring cell with silicone oil filling
Measuring cell with inert liquid
for process temperature -20°C < ϑ
≤ 60°C (-4°F < ϑ ≤ +140°F)
for process temperature 60°C < ϑ
≤ 100°C (max. 85°C for measuring
cell 30 bar) (140°F < ϑ ≤ 212°F
(max. 185°F for measuring cell
435 psi))
SITRANS P DS III/P410 with PROFIBUS PA
Operating Instructions, 02/2016, A5E00053276-08
30 mbar a/3 kPa a/0.44 psi a
30 mbar a + 20 mbar a • (ϑ - 60 °C)/°C
3 kPa a + 2 kPa a • (ϑ - 60 °C)/°C
0.44 psi a + 0.29 psi a • (ϑ - 108 °F)/°F
100 % of max. span (for oxygen measurement: max.
100 bar/10 MPa/1450 psi and 60 °C ambient temperature/process tempera-
ture)
Between the measuring limits (fully adjustable)
Absolute pressure
Span
1)
8.3 ... 250 mbar a
0.83 ... 25 kPa a
3 ... 100 inH
O a
2
43 ... 1300 mbar a
4.3 ... 130 kPa a
17 ... 525 inH
O a
2
160 ... 5000 mbar a
16 ... 500 kPa a
2.32 ... 72.5 psi a
1 ... 30 bar a
0.1 ... 3 MPa a
14.5 ... 435 psi a
5.3 ... 100 bar a
0.5 ... 10 MPa a
76.9 ... 1450 psi a
0 mbar a /kPa a /psi a
30 mbar a /3 kPa a /0.44 psi a
30 mbar a + 20 mbar a • (ϑ - 60 °C)/°C
3 kPa a + 2 kPa a • (ϑ - 60 °C)/°C
0.44 psi a + 0.29 psi a • (ϑ - 108 °F)/°F
Technical data
13.2 SITRANS P DS III input
Maximum
operating pressure MAWP (PS)
32 bar a
3.2 MPa a
464 psi a
32 bar a
3.2 MPa a
464 psi a
32 bar a
3.2 MPa a
464 psi a
160 bar a
16 MPa a
2320 psi a
160 bar a
16 MPa a
2320 psi a
197