The steps to enter your update calibrations are the same as your initial calibration, including
only entering values in one display device. The only difference is, with update calibrations,
you enter your BG meter value just once.
Like the reminders you received with your initial calibration, if you do not enter your BG meter
values right away, the system prompts you every 15 minutes.
Update Calibration Prompts
Device
Smart Device:
Lock Screen
Smart Device:
In App
Receiver
Tap message to clear prompt on your smart device and to clear a prompt on your receiver,
press Select .
Sound/Vibration Prompts
In case you cannot look at your screen, both the smart device and receiver provide, with the
exception of your regular 12 hour update calibration, beep/vibration prompts to let you know
it is time to calibrate or if there was a system calibration error.
For more information on setting your sound/vibration prompts and how to clear them, please
see Chapter 10.
Smart Device
Calibration prompts will alert you with a triple beep if your smart device is not on Silent or Do
Not Disturb .
Dexcom G5 Mobile System User Guide
Calibration
What you see
What it means
Enter update
calibration.
If prompt does not
go away:
• System did
not accept
calibration
• BG values are
very different
from sensor
glucose
readings
What you do
Follow steps in
Sections 6.6 and
6.7.
81