Immunity Test
Voltage Dips, Short
Interruptions and
Voltage Variations
on Power Supply
Input Lines
IEC 61000-4-11
IEC 60601-1-11
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration –
Electromagnetic Immunity
The Dexcom G5 Mobile CGM System is intended for use in the electromagnetic environment specified
in the next table. The customer or the user of the Dexcom G5 Mobile CGM System should ensure that it
is used in such an environment.
Dexcom G5 Mobile System User Guide
Technical Information
(Continued from previous page)
IEC 60601 Test
Level
0% U
for 1 cycle
T
0% U
for 0.5 cycle at
T
8 phase angles
70% U
(30% dip in
T
U
) for 25 cycles
t
0% U
for 250 cycles
T
30 A/m
Transmitter
Compliance
Level
0% U
for 1 cycle
T
0% U
for 0.5 cycle at
T
8 phase angles
70% U
(30% dip in
T
U
) for 25 cycles
t
0% U
for 250 cycles
T
30 A/m
Electromagnetic
Environment Guidance
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.