Table 6-12 Constant Voltage Output Adjustment - Agilent Technologies U3606A User's And Service Manual

Multimeter - dc power supply
Table of Contents

Advertisement

9 Verify the constant voltage adjustments using the

Table 6-12 Constant voltage output adjustment

S2 (8 V/3 A)
S1 (30 V/1 A)
[1] See
CC output adjustment procedure — rear output terminals
Review the
page 141 sections before beginning this procedure.
1 Turn off the instrument and connect the rear panel + and
N O T E
U3606A User's and Service Guide
readback accuracy"
Range
step 8
for information on how to connect the LOAd adjustment item.
"Test Considerations"
– output terminals to the I (red) and LO (black) input terminals.
Ensure that the rear panel sense terminals (+S and –S) are shorted to the
rear panel output (+ and –) terminals. See
connections"
load leads.
on page 157.
Adjustment item
OUt-L
OUt-H
OUt-L
OUt-H
[1]
LOAd
on page 140 and
on page 100 for more information on how to connect the
Calibration Procedures
Adjustments procedures
"CV programming and
Output voltage
0 V
8 V
0 V
30 V
30 V with additional
30 Ω, 50 W load
"Input Connections"
"Remote sensing
6
on
209