Verification And Performance Tests; Recommended Test Equipment; Table 5-1 Recommended Test Equipment For Performance Verification - Agilent Technologies U3606A User's And Service Manual

Multimeter - dc power supply
Table of Contents

Advertisement

5

Verification and Performance Tests

Recommended Test Equipment

Recommended Test Equipment

The test equipment recommended for the performance verification
procedures is listed below. If the exact instrument is not available,
substitute calibration standards of equivalent accuracy.
Table 5-1 Recommended test equipment for performance verification procedures
Application
Recommended equipment
Zero offset and gain verification procedures
Shorting plug — Dual banana plug with
Zero calibration
copper wire short between the two input
terminals
DC voltage
Calibrator
DC current
Calibrator
AC voltage
Calibrator
AC current
Calibrator
Resistance
Calibrator
Function/Arbitrary
Frequency
waveform
generator
Capacitance
Calibrator
Output verification procedures
Constant voltage
programming
Digital multimeter
and readback
Digital multimeter
Constant voltage
load effect (load
Electronic load
regulation)
136
Recommended
accuracy requirements
Fluke 5520A
< 20% of instrument 1 year specification
Fluke 5520A
< 20% of instrument 1 year specification
Fluke 5520A
< 20% of instrument 1 year specification
Fluke 5520A
< 20% of instrument 1 year specification
Fluke 5520A
< 20% of instrument 1 year specification
Agilent 33250A
< 20% of instrument 1 year specification
Fluke 5520A
< 20% of instrument 1 year specification
Agilent 3458A
8½ digit resolution
Agilent 3458A
8½ digit resolution
• Voltage range: 30 V
Agilent 6060B
• Current range: 5 A
• Open and short switches
dc
U3606A User's and Service Guide

Advertisement

Table of Contents
loading

Table of Contents