Table 2-3 Dc Current Measurement Summary - Agilent Technologies U3606A User's And Service Manual

Multimeter - dc power supply
Table of Contents

Advertisement

2
Digital Multimeter Operation
Making Measurements
Measuring DC current
1 Press
DC annunciator

Table 2-3 DC current measurement summary

Available ranges
Measurement method
Burden voltage and shunt
resistance
Input protection
2 Connect the red and black test leads to input terminals I (red) and
3 Probe the test points and read the display.
30
I
to make DC current measurement. (The DC annunciator is
shown on the display.)
Item
LO (black) respectively as shown in
Description
10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
Sigma Delta A-to-D converter
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.1 Ω for 1 A range
• < 0.7 V, 0.01 Ω for 3 A range
Protected with a 3.15 A/500 V, FF fuse
Figure 2-
2.