5
Verification and Performance Tests
Performance Verification Tests
Table 5-5 AC voltage gain verification test
10 mV
100 mV
0.1 V
1 V
1 V
10 V
10 V
100 V
100 V
750 V
C A U T I O N
AC current verification test
1 Connect the calibrator to the front panel I (red) and LO (black) input
2 Press
3 Select each range in the order shown below. Provide the indicated input
154
Input voltage
Input frequency
1 Hz
1 Hz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
1 kHz
Set the calibrator output to 0 V before disconnecting from the
instrument input terminals.
terminals (see
Figure 5- 3
I
to select the AC current function. The AC annunciator will
be illuminated.
current and frequency. Compare the measurement results to the
appropriate test limits shown in
appropriate source settling when using the Fluke 5520A.)
Range
100 mV
100 mV
1 V
1 V
10 V
10 V
100 V
100 V
750 V
750 V
on page 142).
Table 5-
6. (Be certain to allow for
U3606A User's and Service Guide
Error from
nominal 1 year
±0.12 mV
±0.3 mV
±0.0012 V
±0.003 V
±0.012 V
±0.03 V
±0.12 V
±0.3 V
±0.95 V
±2.25 V