Table 6-4 Adjustment Table - Agilent Technologies U1251A User's Manual And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1251A:
Table of Contents

Advertisement

Table 6-4 Adjustment table

Step
Test Function
1
Turn the rotary switch to the
V position
2
Turn the rotary switch to
V position (for model U1252A),
to
V position (for model
U1251A)
Agilent U1251A/U1252A User's and Service Guide
Cal Range
Input
5 V
0.3 V,1 kHz
3 V, 1 kHz
3 V, 50 kHz
50 V
3 V, 1 kHz
30 V, 1 kHz
30 V, 50 kHz
500 V
30 V,1 kHz
300 V,1 kHz
300 V, 50 kHz
1000 V
30 V, 1 kHz
300 V, 1 kHz
300 V, 50 kHz
Short
Dual Banana Plug
with copper wire
short between 2
terminals
5 V
3 V
50 V
30 V
500 V
300 V
1000 V
1000 V
Performance Tests and Calibration
Cal Item
U1251A
0.3000 V
3.0000 V
N/A
03.000 V
30.000 V
N/A
030.00 V
300.00 V
N/A
0030.0 V
0300.0 V
N/A
SHort
3.0000 V
30.000 V
300.00 V
1000.0 V
7
U1252A
0.3000 V
3.0000 V
3.0000 V
03.000 V
30.000 V
30.000 V
030.00 V
300.00 V
300.00 V
0030.0 V
0300.0 V
0300.0 V
SHort
3.0000 V
30.000 V
300.00 V
1000.0 V
137

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1252a

Table of Contents