Agilent Technologies U3606A User's And Service Manual page 233

Multimeter - dc power supply
Table of Contents

Advertisement

30 Ω, 50 W load
5 Verify the constant current adjustments using the
CV output adjustment procedure — rear output terminals
Review the
page 141 sections before beginning this procedure.
1 Turn off the instrument and connect the rear panel + and
N O T E
U3606A User's and Service Guide
+S + + –
– –S
readback accuracy"
"Test Considerations"
– output terminals to the V (red) and LO (black) input terminals.
Ensure that the rear panel sense terminals (+S and –S) are shorted to the
rear panel output (+ and –) terminals. See
connections"
load leads.
on page 163.
on page 140 and
on page 100 for more information on how to connect the
Calibration Procedures
Adjustments procedures
"CC programming and
"Input Connections"
"Remote sensing
6
on
211