Table 2-10 Continuity Function Summary - Agilent Technologies U3606A User's And Service Manual

Multimeter - dc power supply
Table of Contents

Advertisement

2
Digital Multimeter Operation
Making Measurements
Testing continuity
1 Press

Table 2-10 Continuity function summary

Measurement method
Audible tone:
Input protection
2 Connect the red and black test leads to input terminals
3 Probe the test points. The instrument beeps when the continuity
50
again (until the continuity annunciator is shown on the
display) to make continuity measurements.
Item
LO (black) respectively as shown in
measurement is less than or equal to the continuity threshold.
Continuity annunciator
Description
0.83 mA ± 0.2% constant current source, open-circuit voltage limited
to < 5 V
Continuous beep when reading is less than threshold resistance of
10 Ω at 1.0 kΩ range
1000 V
on all ranges, < 0.3 A short circuit
rms
Figure 2-
(red) and
8.
U3606A User's and Service Guide