Agilent Technologies U3606A User's And Service Manual page 135

Multimeter - dc power supply
Table of Contents

Advertisement

Table 4-2 Utility menu item descriptions
Item
Available settings
CV
S1/S2
rAMP
NN.NNN V
rHoLd
t - NNN %
U3606A User's and Service Guide
CC
S1/S2
N.NNNN A
NNNNN S
nH - N.N %
Description
• Press Voltage to select ramp setup for CV.
• Press Current to select ramp setup for CC.
Press Shift >
Range
to toggle between range S1 (30 V/1 A) or
S2 (8 V/3 A)
• The increment of each step will be the amplitude end
position (NN.NNN V/N.NNNN A) divided by the number of
steps (NNNNN).
• The ramp dwelling time will be fastest of output capability
(typically ~300 ms per step).
• See
"Configuring the ramp signal parameters"
for more information.
• Set the refresh hold variation (t - NNN %) in percentage.
When the variation of the measuring value exceeds the
preset setting, the refresh hold will be ready to trigger.
• Set variation to "OFF" to enable data hold mode.
• Set the refresh hold threshold (nH - N.N %) in percentage
for voltage, current, and capacitance measurements. The
reading value will not be updated when the reading falls
below the threshold value.
• See
"Enable refresh hold"
hold"
on page 127 for more information.
System Related Operation
Utility Menu Summary
on page 122
on page 126 and
"Enable data