4
System Related Operation
Utility Menu Summary
Table 4-2 Utility menu item descriptions
Item
Available settings
GPib
IOb
NN
SELF.t
no
NNNN Ω
db.rEF
CV
S1/S2
SCAn
NN.NNN V
112
U-CdC
U-tMC
YES
CC
S1/S2
N.NNNN A
NNN-NN s
Description
• Select "GPib", "U-tMC", or "U-CdC" as the desired remote
interface connection.
• "USB-tMC" simulates the USB interface according to
USB-TMC standard.
• "USB-CdC" is used to simulate the communication port.
• See
"Connecting to a remote interface"
"Remote Operation"
on page 130 for more information.
• Set the GPIB address from 1 to 30.
• See
"Configuring and connecting the GPIB interface"
page 131 for more information.
• If "YES" is selected, the instrument exits the utility menu
and immediately executes the self-test.
• After the self-test completes, the instrument returns to
normal operation.
• See
"Performing a self-test"
information.
• Set dB reference impedance value from 1 Ω to 9999 Ω.
• See
"Selecting a dBm reference resistance value"
page 121 for more information.
• Press Voltage to select scan setup for CV.
• Press Current to select scan setup for CC.
Press Shift >
Range
to toggle between range S1 (30 V/1 A) or
S2 (8 V/3 A)
• The increment of each step will be the amplitude end
position (NN.NNN V/N.NNNN A) divided by the number of
steps (NNN S).
• Set the scan dwelling time (NN s) from 1 s to 99 s.
• See
"Configuring the scan signal parameters"
for more information.