CC output adjustment procedure — front output terminals
Review the
page 141 sections before beginning this procedure.
1 Turn off the instrument and connect the front panel
2 Turn on the instrument. Press Current to enter the constant current
3 The primary display will show the uncalibrated value and the upper
4 Configure each adjustment item shown in
5 Use
N O T E
6 Press
U3606A User's and Service Guide
"Test Considerations"
(black) output terminals to the I (red) and LO (black) input
terminals.
output calibration. Ensure that the S2 (8 V/3 A) range is selected (the
S2 annunciator is illuminated). Press Current again if it is not.
secondary display will show the reference level of the adjustment item.
or
to select the adjustment item.
Always complete tests in the same order as shown in
O U T
to start the adjustment. The display "CALib" in the lower
S BY
secondary display starts flashing to indicate that the calibration is in
progress.
• Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing "PASS".
Calibration Procedures
Adjustments procedures
on page 140 and
"Input Connections"
Table 6-
on
(red) and
11.
Table
6-11.