7 Press
8 Repeat
9 Verify the frequency gain adjustments using the

Table 6-5 Frequency gain adjustment

1000.00 Hz, 1 V
Resistance gain adjustment procedure
Review the
considerations"
This procedure adjusts the gain for the two- wire resistance function. The
gain for the 100 MΩ range is derived from the 10 MΩ range and does not
have a separate adjustment point.
1 Press
2 The primary display will show the uncalibrated value and the upper
3 Configure each adjustment item shown in
N O T E
U3606A User's and Service Guide
O U T
to start the adjustment. The display "CALib" in the lower
S BY
secondary display starts flashing to indicate that the calibration is in
progress.
• Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing "PASS".
• An adjustment failure is indicated by a long beep, the primary
display showing "FAiL" and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
step 3
through
Table 6-
5.
verification test"
Adjustment item
rms
"Test Considerations"
on page 191 sections before beginning this procedure.
to enter the resistance gain calibration.
secondary display will show the reference value of the adjustment item.
If the zero offset adjustment procedure has been recently performed prior
to the resistance gain calibration procedure, the adjustment items "Short"
and "OPEn" can be omitted.
step 7
for each gain adjustment point shown in
on page 156.
Input frequency
1 kHz
on page 140 and
Calibration Procedures
Adjustments procedures
"Frequency gain
Input voltage (V
rms
1 V
rms
"Gain adjustment
Table 6-
6.
6
)
197