Calibration Process - Agilent Technologies U3606A User's And Service Manual

Multimeter - dc power supply
Table of Contents

Advertisement

Calibration Process

The test equipment recommended for the calibration process and
procedure is listed in
The following general procedure is the recommended method to complete
a full instrument calibration.
1 Read the
2 Perform the verification tests to characterize the instrument (See
3 Unsecure the instrument for calibration. (See
4 Perform the adjustment procedures
5 Secure the instrument against calibration. Change the calibration code if
6 Note the new security code (if it has been changed) and calibration
N O T E
U3606A User's and Service Guide
"Test Considerations"
"Verification and Performance Tests Overview"
instrument for calibration"
page 189).
necessary. (See
"Changing the calibration security code"
count in the instrument maintenance records.
Ensure that you quit the adjustment mode before powering off the
instrument.
"Recommended Test Equipment"
on page 140.
on page 176.)
("Adjustments procedures"
Calibration Procedures

Calibration Process

on page 174.
on page 148).
"Unsecuring the
on
on page 179.)