Calibration Procedures
Adjustments procedures
N O T E
3 Press Shift > Hold. The display "CALib" in the lower secondary display
4 The primary display will show the measurement functions and ranges
5 Remove the shorting plug from the input terminals.
6 Press Shift > MinMax. The display "CALib" in the lower secondary
7 The primary display will show the measurement functions and ranges
8 Perform the
190
To minimize thermal effects, wait at least 1 minute after connecting the
shorting plug before executing the zero offset adjustment.
starts flashing to indicate that the calibration is in progress.
as the adjustments progress.
• Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing "PASS".
• An adjustment failure is indicated by a long beep, the primary
display showing "FAiL" and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
display starts flashing to indicate that the calibration is in progress.
as the adjustments progress.
• Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing "PASS".
• An adjustment failure is indicated by a long beep, the primary
display showing "FAiL" and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
"Zero offset verification test"
calibration results.
on page 150 to check the zero