N O T E
U3606A User's and Service Guide
• The OV protection feature is always enabled by default whenever you
select constant current mode. You cannot disable the OV protection.
• If the OV value is set to a greater value than the OVP value, the OVP
value will be adjusted to equal the OV value.
• If the OV value is set to zero, this will result in the output current being
dropped to zero for limiting.
• The OV value is not applicable for square-wave output.
• The OV value is limited by the range selected. Press Shift >
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).
DC Power Supply Operation
Protection Functions
Range
3
to
85