Performing Voltage Measurements
Ensure that the terminal connections are connected correctly before making any
measurement. To avoid damaging the multimeter, do not exceed the rated input limit.
Measuring AC Voltage
• Five ranges:
• Measurement method:
• Crest factor:
• Input impedance:
• Input protection:
1 Press
.
2 Connect the red and black test leads to the respective input terminals as
shown in
Figure 2-
3 Probe the test points and read the display.
In autoranging mode, the multimeter automatically selects the appropriate
range and the measurement is displayed.
+
ACV voltage
source
–
• Slow reading rate: 120.000 mV, 1.20000 V, 12.0000 V,
120.000 V, 750.00 V.
• Medium reading rate: 400.00 mV, 4.0000 V, 40.000 V,
400.00 V, 750.0 V.
• Fast reading rate: 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 750 V.
AC coupled true RMS—measures the AC component with up to
400 VDC bias on any range
Maximum 3:0 at full scale
1 M ±2% in parallel with <120 pF on all ranges
750 V RMS on all ranges