Table 5-2 Zero offset verification test

Short
Open
U3606A User's and Service Guide
Input
Function
DC voltage
Resistance
DC current
Verification and Performance Tests
Performance Verification Tests
Error from
Range
nominal 1 year
100 mV
±0.008 mV
1 V
±0.00005 V
10 V
±0.0005 V
100 V
±0.005 V
1000 V
±0.05 V
100 Ω
±0.008 Ω
1 kΩ
±0.00005 kΩ
10 kΩ
±0.0005 kΩ
100 kΩ
±0.005 kΩ
1 MΩ
±0.00005 MΩ
10 MΩ
±0.0005 MΩ
100 MΩ
±0.005 MΩ
10 mA
±0.0015 mA
100 mA
±0.005 mA
1 A
±0.00007 A
3 A
±0.00021 A
5