DC Power Supply Operation
Protection Functions

Overvoltage limit (OV)

The overvoltage limit prevents the output voltage present across the load
to change beyond the programmed overvoltage limit. If the load effect
exceeds the programmed overvoltage limit, the CC output will be lowered
to maintain the output power across the load. The combination of the OV
and OVP protection functions create a closed loop circuit protection for
sensitive load behaviors.
Setting the OV level
The OV level can be programmed when the output is enabled (OUT) or
disabled (SBY). Ensure that the U3606A is operating in constant current
mode. (The CC annunciator is illuminated.)
1 Press Shift >
2 Press
3 Press
4 Press Shift >
5 Alternatively, press Shift >
84
Limit
to enter the OV edit menu.
or
to select which digit position to be changed.
or
to increase or decrease the value. The OV value is set
to the maximum protection limits by default.
• 30.000 V when range S1 is selected, and
• 8.000 V when range S2 is selected.
Save
or Current (Limit) to save the changes. "donE" will be
briefly displayed in the upper secondary display. The display will then
return to normal.
without saving.
Exit
or Shift >
Limit
U3606A User's and Service Guide
to exit the edit mode