Calibration Message

The instrument allows you to store a message in calibration memory. For
example, you can store information such as the date when the last
calibration was performed, the date when the next calibration is due, the
instrument serial number, or even the name and phone number of the
person to contact for a new calibration. The calibration message may
contain up to 40 characters.
• You can record a calibration message only when the instrument is
• You can read the calibration message (from the remote interface only)
To store a calibration message use the CALibration:STRing and
CALibration:STRing? commands from the remote interface.
& CAL:STR "CAL: 27 Nov 2009"
& CAL:STR?
$ "CAL: 27 Nov 2009"
See the U3606A Programmer's Reference for more information on the
CALibration subsystem.
U3606A User's and Service Guide
unsecured.
whether the instrument is secured or unsecured.
Calibration Procedures

Calibration Message

This command stores a message in the
calibration memory.
This query returns the message currently stored
in calibration memory (the quotes are also
returned).