Configuring the scan signal parameters

The following procedure shows you how to configure the scan signal
parameters.

Table 4-4 Scan signal parameters

Scan signal
Items
Range
Amplitude end position
Number of steps
Dwelling time
1 To access the utility menu, press Shift > Utility.
2 The first utility menu item (Error) will be shown in the upper
3 Press
4 Press Voltage or Current if you wish to configure the scan signal
5 Select an appropriate output range by pressing Shift > Range. The scan
124
0 V to 31.500 V
secondary display.
or
until the menu item "SCAn" is shown in the upper
secondary display.
parameters for CV or CC output. (The CV or CC annunciator
illuminates respectively when selected.)
signal amplitude end position is limited by the range selected. (The S1
or S2 annunciator illuminates respectively when selected.)
Output
Constant Voltage
S1
S2
0 V to 8.400 V
1 step to 100 steps
1 s to 99 s
Constant Current
S1
0 A to 1.0500 A
0 A to 3.1500 A
U3606A User's and Service Guide