5
Verification and Performance Tests
Additional Verification Tests
Table 5-15 Optional capacitance gain verification test
1000 μF
10000 μF
N O T E

Optional square-wave output verification test

This test verifies that the square- wave amplitude, frequency, and duty
cycle functions are within specifications. Note that the measurement
values over the remote interface should be identical to those displayed on
the front panel.
N O T E
Square-wave amplitude output verification test
1 Turn off the instrument and connect an oscilloscope between the front
2 Turn on the instrument. Press
3 Output each value in the order shown below. Record the output voltage
4 Repeat
168
Input capacitance
For the best accuracy, take a zero null measurement with open test leads
to null out the test lead capacitance, before connecting the test leads to
the calibrator.
You should consider programming the U3606A over a remote interface for
this test to avoid round off errors. See the U3606A Programmer's
Reference for more information on remote interface programming.
panel
(red) and
page 147).
Ensure that the S1 (30 V/1 A) range is selected (the S1 annunciator is
illuminated). Press Shift >
reading on the oscilloscope. Compare the measurement results to the
appropriate test limits shown in
step 2
and
to select the S2 range (the S2 annunciator is illuminated).
Range
1000 μF
10000 μF
(black) output terminals (see
to select the square- wave mode.
Range
again if it is not.
Table 5-
for the S2 (8 V/3 A) range. Press Shift >
step 3
Error from nominal 1 year
±15 μF
±250μF
Figure 5- 12
16.
U3606A User's and Service Guide
on
Range