Performance Tests and Calibration
Table 14
Valid Gain Adjustment Input Values
Function
DC Volts
DC Current
Ohms
Frequency
AC Current
AC Volts
Capacitance

DC Voltage Gain Adjustment Procedure

N O TE
96
Range
100 mV, 1 V, 10 V, 100 V, 1000 V
10 mA, 100 mA, 1000 mA, 10 A
100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1MΩ, 10 M Ω 0.9 to 1.1 x Full Scale
Autorange/1 kHz
1 mA, 10 mA, 100 mA, 1000 mA, 10 A
10 mV, 100 mV, 1 V, 10 V, 100 V, 750 V
0.4 nF, 1 nF, 10 nF, 100 nF, 1 µF, 10 µF,
100 µF, 1000 µF, 10000 µF
Review the
"Test Considerations"
Adjustment Considerations"
beginning this procedure.
1 Press
D C V
2 The primary display will show the uncalibrated value and
the secondary display will show the reference value of the
Cal Item.
3 Configure each Cal Item shown in the adjustment table
below.
If the zero adjustment procedure has been recently performed prior to the
DC Voltage gain calibration procedure, the Cal Item 'Short' can be
neglected.
4 Use
Item.
Valid Amplitude Input Values
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
Input
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
on page 71 and
on page 95 sections before
to enter DC Voltage Gain Calibration.
(Auto) or
34405A User's and Service Guide
100 mV rms, 900 Hz to 1100 Hz
"Gain
(Range) to select the Cal