Table 6-2 Verification Test - Agilent Technologies U1251B User's And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1251B:
Table of Contents

Advertisement

Table 6-2 Verification Test

Step
Test Function
1
Turn the rotary switch to the
position
2
Press
frequency mode
3
Press
Duty Cycle mode
Agilent U1251B/U1252B User's and Service Guide
Perform the verification test steps in the following
5 V
[1]
50 V
500 V
1000 V
button to go to
9.9999 kHz
button to go to
0.01% – 99.99%
Range
5520A Output
5 V, 1 kHz
4.5 V, 10 kHz
4.5 V, 20 kHz
4.5 V, 30 kHz
4.5 V,100 kHz
50 V,1 kHz
45 V,10 kHz
45 V, 20 kHz
45 V, 30 kHz
45 V, 100 kHz
500 V, 1 kHz
1000 V, 1 kHz
0.48 V, 1 kHz
5.0 Vpp @
50%, Square
Wave, 50 Hz
Performance Tests and Calibration
Error from nominal 1 year
U1251B
± 32.5 mV
± 169.5 mV
N/A
± 169.5 mV
N/A
± 325 mV
± 1.695 V
N/A
± 1.695 V
N/A
± 3.25 V
± 10 V
± 500 mHz
± 0.315%
6
Table 6-
2:
U1252B
± 22.5 mV
± 71.5 mV
± 169.5 mV
N/A
± 169.5 mV
± 225 mV
± 715 mV
± 1.695 V
N/A
± 1.695 V
± 2.25 V
± 8.0 V
± 500 mHz
± 0.315%
129