Access Level For Disabled Logging - RADWAG WPW/E User Manual

Table of Contents

Advertisement

Notice:
This setting is valid although standard logging is disabled.

18.2.2. Access level for disabled logging

The scale program gives possibility to set an access level in case
of disabling logging procedure.
Procedure:
• Enter <P6 DATABASES> and then:
OPERATOR
ADVANCED OP.
ADMIN
Return to weighing:
See 11.3.
-
Can perform weighings, edit P1, P4, P9
parameters; cannot edit databases and erase
statistics; cannot change weighing thresholds
Min, Max, cannot adjust the standard mass
in modes "counting pieces" and "deviations".
-
Can perform weighings and edit parameters
except P8, P10 and P6.1, P6.2, P6.3; can
edit databases except operators;
-
Have a full access to parameters and
databases, can perform user calibration
(in non-verified scales).