n
- weighing platform number
mass - 9 characters justified to the right
unit - 3 characters justified to the left
Example:
Let us assume that both platforms are connected to indicator PUE C41H.
S I A CR LF – computer command
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF - command done, mass values
from both platforms are returned in basic units

28.4.8. Send the stable result in current unit

Syntax: SU CR LF
Possible answers:
SU_A CR LF
SU_E CR LF
SU_I CR LF
SU_A CR LF
MASS FRAME
Frame format:
1
2
3
S
U
space
Example:
S U CR LF – computer command
S U _ A CR LF - command accepted and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass
value in current unit is returned.
- command accepted and in progress
- timeout while waiting for stable results
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in current unit is returned
4
5
stability
space
6
7-15
16
sign
mass
space
- 160 -
17
18
19
20
unit
CR
21
LF