27.5.1. Technical specification

Parameters of outputs
Quantity of outputs
Type of outputs
Wire diameter
Maximal load-current contact capacity
Maximal forward voltage
Parameters of inputs
Quantity of inputs
Input type
Wire diameter
Control voltage range

27.5.2. Installing method in PUE C41H indicators

Module is designated for assembly inside indicator PUE C41H. Module
is assembled to main board of indicator to 10-pin J12 connector. For
module WE8 gland is installed on casing lid, 3m cable ended with
insulated conductors is led through gland.
Installing procedure:
1. Unplug the indicator (remove the plug from socket 230V);
2. Unscrew the casing lid (back part of indicator casing);
3. Install the module in connector J12 on main board;
4. While installing module pay attention to plastic posts fastening
to main board. They should be placed in assembly holes in main
board and in assembly holes in module WE 8,
5. In one of free glands remove the stopper and led through it cable
IN/OUT (if necessary use bigger gland depending on cable diameter);
6. Connect the cable IN/OUT to joint J1, J2 for outputs and J3, J4 for
inputs on module 8IN/8OUT according to description given in table;
7. Connect cable IN/OUT to group of conductors (undo band clips
fastening group of conductors, lay the cable and clamp band clips –
multiple use band clips);
8. Assembly cover of indicator casing.
8
Reed operation contacts
0,14 - 0,5mm
0,2A DC
50V DC
8
Optoinsulated
0,14 – 0,5mm
5 -24V DC
- 145 -