Example 1:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – the printout generated from
the scale after pressing ENTER/PRINT.
Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.
Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.

28.6. Continuous transmission

The indicator can work in a continuous transmission mode. It can be
switched on or off in parameters or using RS232 commands.
Frame format sent by the indicator when continuous transmission is set:
1
stability
space
Stability character
sign
mass
unit
command

28.7. Configuring printouts

If some information included are redundant or not sufficient and there is
a necessity of changes you can design a non-standard printout. There is
possible to create up to four standard printouts (see the chapter. 17).
2
3
4 -12
sign
mass
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range
[space] for positive values or
[-] for negative values
9 characters justified to the right
3 characters justified to the left
3 characters justified to the left
13
14
space
- 166 -
15
16
17
unit
CR
18
LF