29. ERROR MESSAGES

ERROR "XXX" ESC – RETURN TO PREVIOUS SETTING
(where: XXX – parameter name) – confirmed with ENTER
of a wrong value in the user's menu,
-
Value beyond the zero range,
Err2
-
Value beyond the tare range,
Err3
-
Calibration mass or start mass adjustment error
Err4
(±1% for weight, ±10% for start mass),
-
Mass of a single piece lower than the 0.1 of scale division,
Err5
-
Mass of all pcs (declared as standard) lower than one
Err6
division,
-
Tarring / zeroing operation time exceeded,
Err8
-
Zero value from the AD converter,
NULL
FULL2
-
Measurement range overflow,
HI
-
Display range overflow,
-
Start mass error, the mass on the weighing platform
LH
is beyond the acceptable range ±20 of start mass

30. TROUBLE SHOOTING

Problem
Scale switches off
automatically
„LH" appears on the
display
Cause
„t1" parameter set to „YES"
(Power save)
Weight pan not empty
when switching on
- 167 -
Solution
In <P9 OTHERS> change
<P9.3 AUTO SWITCH.> to
„NO"
Unload the pan. Indication