23.1. Calibration procedure

• Enter <P10 USER CALIB.> according to 11.2. and then:
• Unload the pan,
• Press
ADJ. START MASS, in the bottom line,
• After completing this procedure you will see the following window:
• Place the required mass on the platform 1 then press
• During the calibration process you will see: ADJ. CALIB. FACTOR,
in the bottom line.,
• After the procedure is completed you will see in: UNLOAD THE PAN
in the bottom line,
• After taking off the calibration weight :
• Return to weighing, saving parameters.
Return to weighing:
See 11.3.
. During adjusting a start mass you will see: