Technical Specification; The Way Of Installing Inside Pue C41H - RADWAG WPW/E User Manual

Table of Contents

Advertisement

Module accessible in three configurations:
Voltage output AN 0-10V
Current output AN 4-20mA
Current output AN 0-20mA

27.2.1. Technical specification

Work modes
Resolution
Current output resistance
Voltage output resistance
Power supply

27.2.2. The way of installing inside PUE C41H

These modules are intended to mount inside PUE C41H. They need to be
connected to the 10-pin J12 connector. For all configurations of AN, there
is a gland installed on the back wall of the housing. A 3-meter shielded
cables are led out via the gland. Wires should be free from insulation.
Installing procedure:
1. Unplug the scale from mains;
2. Unscrew and take off the back wall of the housing;
3. Install your module in J12 on the main board;
4. During installation turn your attention to plastic columns. They
should be placed one side in mounting holes in the main board
and the other side in the mounted module;
5. Led the PT0015 cable through one of the free glands;
6. Connect the PT0015 cable to J3 on the analogue module according
to the description below;
7. Connect the PT0015 cable shield to the housing
(screwed terminator, 4mm diameter);
8. The cable connect to the group of wires (unhook band clips
fastening the group of wires, lay the cable and hook the band clips).
Band clips of multiple usage;
9. Screw down the back wall.
4 - 20mA , 0 - 20mA, 0 - 10V
16 bit
<500
>400
24V DC (12 - 30V DC) max 40mA