fig. 27
Replacing the control panel
6.7.2 Replacing the Control Panel Holder
The control panel holder for the A200 power wheelchair is clamped to the underside of the arm
rest with an iron rail.
Preparation:
❒ Remove the side panel.
❒ Remove and set aside the control panel.
Tools:
❒ Allen wrench, 3 mm
Steps:
1. Remove the side panel from the holder.
2. Loosen the two set screws.
3. Change the position of the control panel holder, or remove the control panel holder and re-
place if required.
fig. 28
Replacing the control panel holder
31