Remove the Existing Lamp

1. Tap and hold the red power button
off.
2. Allow the lamp to cool for a minimum of 15 minutes.
U
3.
nplug the projector.
4. Put on your protective clothing and face shield.
5. Insert the key in the lock on the lamp door, turn the key, and then open the lamp door. Do not
place heavy objects on the open lamp door.
6. Loosen the two thumbscrews (A and B) and open the lamp access door.
Remove the old lamp and inspect the reflector
7.
a. Remove the screw securing the anode wire (B).
b. Loosen the cathode screw (D) on the cathode clamp (F).
c. Hold the lamp from the anode end and carefully unscrew and remove the cathode nut (E).
d. Hold the lamp from the anode end and carefully slide out ensuring not to make contact
with the reflector.
e. With your free hand guide the cathode end out of the reflector, on an angle through the
lamp access door.
f. Before placing the old lamp into the protective case make sure the cathode nut is
reinstalled. Place the lamp, within the case, on the floor where it cannot fall or be bumped.
Handle the box with extreme caution - the lamp is hazardous even when
packaged. Dispose of lamp box according to local area safety regulations.
Failure to comply could result in death or serious injury.
CP2215 User Manual
020-101225-01 Rev. 1 (01-2014)
on the TPC Main panel to turn the lamp and projector
:
Maintenance
79