Chapter 4
Patient stations
60
Pressing Cancel for 8 seconds (and 4 feedback tones) on a
patient station or associated room station clears all calls made
from the room or bed, and all reminders for those calls. Pressing
Cancel for 8 seconds (and 4 feedback tones) also resets all
stations in the room.

Placing the patient station in privacy mode

Privacy mode blocks audio connections. Placing a call from the
patient station, overrides privacy mode.
To turn privacy mode on or off:
1.
Click the Privacy button on the patient station.
The Status light slowly flashes green and the Privacy light turns
on, indicating privacy mode is in effect.
Until privacy mode is canceled, or a higher priority event that
uses the status light occurs, the status light will continue to
slowly flash green.
2.
Click the Privacy button again to cancel privacy mode.

Placing the patient station in staff follow mode

Staff follow modes at the patient station
During system installation, patient stations can be configured with
different staff follow modes. Possible staff follow modes for a patient
station are:
No staff follow made, no calls annunciated.
Duty area mode: The patient station annunciates calls
originated in its duty area.
Console mode: The patient station annunciates calls originated
in its assigned staff console's coverage area, which may consist
of multiple duty areas, or possibly one or more nursing units.
When a patient station is configured for duty area mode or console
mode, it enters this mode and briefly annunciates the highest
priority call originated in the corresponding coverage are at the time
that a staff member registers presence in the room. New nurse calls
may be briefly annunciated until presence is no longer registered.
For information on configuring the system's staff follow mode, see
the installation manual provided with your Telligence system
3101010 R1.0 Telligence Patient-Staff Communications System
Installation Manual
Telligence Patient-Staff Communication System Operation Manual