Telligence Patient-Staff Communication System Operation Manual
stations used only by staff, see Chapter 5 "Staff/duty stations"
starting on page 89.
Call cord stations
The call cord on the HC-CCSTN call cord station provides patients
with a simple means of placing a nurse call. Depending on station
configuration, pressing the button at the end of the station's call
cord places a nurse call with a priority of either patient normal,
patient priority, or patient emergency.
Figure 31: Call cord station
A call cord station may be associated with other stations; or it may
be a standalone station with its own identifiable location (for
instance, as the sole station in an exam room). Call cord stations
have no audio capability and cannot be used to communicate with
staff members at staff consoles or annunciators.
1.
Press the button at the end of this station's call cord to place a
nurse call. When you do so, the following happens:
The Call lights on any associated patient or staff/duty
•
station turn on.
The call assurance tone sounds at any associated patient
•
or staff/duty station.
Associated corridor and zone lights turn on according to
•
the call cord station's configured priority:
For patient normal, dome lights show a steady white light.
•
For patient priority, dome lights slowly flash an amber light.
•
For patient emergency, dome lights fast flash an amber light.
•
Tones sound at staff consoles and annunciators, and at
•
stations where staff follow is active.
Chapter 4
Patient stations
81
Need help?
Do you have a question about the Telligence Patient-Staff Communications System and is the answer not in the manual?
Questions and answers