Chapter 5
Staff/duty stations
106
2.
To cancel the staff emergency call, press the Cancel button on
the originating staff emergency station, or on any associated
patient or staff room station.
•
Emerg button lights extinguish.
•
Any associated patient or staff/duty station's Call lights
extinguish.
•
Dome lights extinguish.
•
Tones stop sounding at staff consoles and annunciators,
and at stations where staff follow is active.
•
The nurse call row disappears from staff consoles and
annunciators.
If the staff emergency call had a reminder set for it, the
following happens when the Cancel button is pressed:
•
Dome light display of the staff level reminder ceases.
•
The reminder row disappears from staff console or
annunciator Reminders screens.
Call stations with smoke input
The model HC-PB2-CALLIN call station is a generic nurse call station.
This two-button station may be equipped with a pull cord, can be a
standalone station, or can be associated with a patient or staff/duty
station. The station may be installed in a patient room or other
rooms where patients are temporarily located, such as exam or
recovery rooms, or it may be located in various staff rooms. A call
station has no audio capability, and cannot be used to
communicate with staff members at staff consoles or annunciators.
Figure 40: Call station with smoke input (shown without pull cord)
The station, which normally places a staff normal nurse call, has an
input that can be connected to a smoke detector. If connected to a
Telligence Patient-Staff Communication System Operation Manual