To Operate The Function Menu; To Perform Sampling Setup - Owon PDS5022S User Manual

Pds series portable colour digital storage oscilloscope
Hide thumbs Also See for PDS5022S:
Table of Contents

Advertisement

User's Manual of OWON Colour Digital Storage Oscilloscope
this Holdoff period,the trigger system becomes "blind " to grigger signals.This function
helps to view complex signals such as an AM waveform.Press Holdoff button to activate
"TRIG LEVEL" knob,then turn it to adjust Holdoff time.

To operate the Function Menu

The function menu control zone includes 6 function menu buttons and 3
immediate-execution buttons: SAVE/RCL, MEASURE, ACQUIRE, UTILITY,
CURSOR, DISPLAY,AUTOSCALE, AUTOSET and RUN/STOP.

To perform Sampling Setup

Press the ACQUIRE button and the menu is displayed in the screen, shown as Fig.43.
The description of the Sampling Setup Menu is shown as follows:
Function Menu
Sample
Peak Detect
Average
Averages
Fig.43 ACQU MODE Menu
Setting
Description
General sampling mode.
It is used for the detection of the jamming burr
and the possibility of reducing the confusion.
It is used to reduce the random and don't-care
noises, with the optional number of averages.
4, 16, 64, 128
Choose the number of averages.