User's Manual of OWON Colour Digital Storage Oscilloscope
Zoom in
Press the F3 menu selection button and choose Zone Window. As a result, the
window area defined by two cursors will be expanded to the full screen size (see
Fig.29).

To set the Trigger System

When the oscilloscope begins to collect the data and display the wave form depends
on a trigger. Once it is set correctly, the trigger can transfer the unstable display into a
meaningful wave form.
When beginning to collect data, the oscilloscope will collect adequate data to draw the
wave form at the left side of the trigger point at first. It will continuously perform the
data acquisition while waiting for the trigger condition. After a trigger is detected, the
oscilloscope will continuously collect data enough to draw the wave form at the right
side of the trigger point.
One knob and four function menu buttons are included in the trigger control zone.
TRIG LEVEL: Trigger the level control knob and set the signal voltage
corresponding to the trigger point.
SET TO 50%: Set the trigger level as the vertical mid point value of the amplitude of
the trigger signal.
FORCE TRIG: It is a force trigger button for the generation of a trigger signal,
Fig.29 Zone Window
39