Example 3: Capture The Single Signal - Owon PDS5022S User Manual

Pds series portable colour digital storage oscilloscope
Hide thumbs Also See for PDS5022S:
Table of Contents

Advertisement

User's Manual of OWON Colour Digital Storage Oscilloscope
menu (see Fig.65).
10. Calculate the amplifier gain with the following formulas.
Gain = Output Signal / Input signal
Gain (db) = 20×log (gain)

Example 3: Capture the Single Signal

The digital storage oscilloscope takes the lead in providing the convenience capturing
of such non-periodic signals as pulse and burr, etc. If you intent to capture a single
signal, you can not set the trigger level and the trigger edge unless you have particular
priori knowledge of this signal. For example, if the pulse is the logic signal of a TTL
level, the trigger level should be set to 2 volts and the trigger edge be set as the rising
edge trigger. If it is uncertain as to the signal, you can make an observation of it in
advance under the automatic or ordinary mode to determine the trigger level and the
trigger edge.
The operation steps are as follows: