Screen Reference
CONFIGURATION
156
Input Field
a
Specifies one traffic channel depending on the parameter selection
TCH (Talk)
in the Variable field.
If TCH is selected in the Variable field, this specifies one traffic
channel to make Talk tests in addition to six traffic channels for RF
tests.
If DC Power is selected in the Variable field, this specifies one
traffic channel to make both Talk tests and RF tests.
The frequency corresponding to the channel accompanies. The
allowable range is as follows:
GSM900: 1 to 124
E-GSM: 0 to 124, and 975 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810
Specifies to either TCH or DC Power as follows to make up to six
Variable
measurements:
o TCH: Enables to make six measurements on six traffic
channels specified in the TCH input fields of the Test
Item table, with or without one fixed dc voltage.
o DC Power: Enables to make six measurements on one
traffic channel with six dc voltages specified in the DCP
input fields of the Test Item table.
According to the selection of Variable field, either of the following
TCH or DCP
parameters need to be specified for a new test:
o TCH: Accepts up to six traffic channels. The frequency
corresponding to the channel accompanies. The
allowable range is as follows:
o DCP: Accepts up to six dc voltages ranging from 3.0 to
11.0 V in 0.1 V steps.
a. The variable increment softkeys are activated. Refer to
ment Softkey" on page
b. When DC Power is set to Off, the DC Current cannot be measured and
"----" is shown in the DC Current cells in the table of the AUTO-
MATIC TEST: Stand-by detailed screen.
Description
GSM900: 1 to 124
E-GSM: 0 to 124, and 975 to 1023
DCS1800: 512 to 885
PCS1900: 512 to 810