T
Error Code E11
ECMOLIFE
T
Disconected
58
Internal Battery Failure
ECMOLIFE device has been
disconnected, the treatment control
switch to Colibrì device
The fuse of the internal battery has been detected as broken.
From now on, if you disconnect the power supply corde, the device
High
will be shut down.
Turn off the device and switch to back-up section.
Contact Technical Assistance.
The connection between ECMOLIFE system and COLIBRI system
has been interrupt. If the connection cable has been detached, re-
connect it to the console. If the "Disconnect COLIBRI" softkey ON
High
ECMOLIFE has been pressed, reconnect the systems through the
"Connect COLIBRI" softkey. If the connection fails, continue the
treatment on COLIBRI device.