5.3. FUNCTIONALITIES

TURNING ON THE DEVICE
Once COLIBRÌ Console is turned on, the diagnostic page will be displayed.
DIAGNOSTIC PHASE
During the diagnostic phase, the device will check if
motor driver and flowmeter are properly connected.
Be sure the centrifugal pump is inserted and correctly
primed during the diagnostic phase.
In order to safely proceed with the operations,
acoustic Audio Alarm detection shall be confirmed
manually.
After conformation of Audio Alarm detection, you can
switch second diagnostic page in which the device will
check if pressure sensors are properly connected.
Each subsequent connection of pressure sensors in this phase or during the treatment shall be confirmed in the
diagnostic page by clicking the Retry button.
All undetected probes or sensors won't sample parameters.
Warning
Do not continue if acoustic alarm is not detected at the end of this phase. ALARM SYSTEM might
be compromised.