SPI Master
T
Motor Timeout
SW ERROR
T
(LPSM)
T
Motor Fault
T
Error Code E03
T
Error Code E05
T
Error Code E06
T
Error Code E07
T
Error Code E08
T
Error Code E09
T
Error Code E10
en
A technical error is detected on Motor
board
A SW error is detected
An expected error occurs on Motor board
CRC Error
Reset Master board
Unexpected reset
UnexpectedPower reset
Reset motor failure
Internal Battery Error
Slave SW version fail
An audible alarm will sound and the System will continue to
operate. Press the alarm acknowledge button, if the visual alarm
message does not disappear, clamp the return tubing, stop the
High
Pump and switch to the backup COLIBRI', Motor and Flowmeter.
Resume support. Record the alarm message and contact Technical
Assistance.
An audible alarm will sound and the System will continue to
operate. Press the alarm acknowledge button, if the visual alarm
message does not disappear, clamp the return tubing, stop the
High
Pump and switch to the backup COLIBRI', Motor and Flowmeter.
Resume support. Record the alarm message and contact Technical
Assistance.
Press the alarm acknowledge button, if the message does not
disappear, clamp the return tubing, stop the Pump and switch to
High
the backup COLIBRI', Motor and Flowmeter. Resume support.
Record the fault code message and contact Technical Assistance.
This error might be caused by a SW update. Restart the device.
High
If the problem persists please record the fault code message and
contact Technical Assistance.
Clamp the return line and switch to back-up section to proceed the
High
treatment. Record the fault code message and contact Technical
Assistance
Clamp the return line and switch to back-up section to proceed the
High
treatment. Record the fault code message and contact Technical
Assistance.
This error might be caused by an unexpected power interruption due
to battery discharge or emergency button activation.
High
Normal operation can be restored by turning off and on the device.
If none of those circumstances occurred, record the fault code
message and contact Technical Assistance.
Clamp the return line and switch to back-up section to proceed the
High
treatment. Record the fault code message and contact Technical
Assistance.
This error might be displayed if, during the previous treatment, the
Internal battery fault error is activated. Restart the device.
High
If the problem persists please record the fault code message and
contact Technical Assistance.
This error might be caused by a incorrect SW update. Restart the
device.
High
If the problem persists please record the fault code message and
contact Technical Assistance.
57