At the end of the diagnostic phase, by pressing the start treatment button the device is able to check the type of
flow probe connected to the device.
A warning screen will then appear where it will be necessary to confirm the type of treatment to be carried out
(Adult/Paediatric or New Born) compatible with the type of flowmeter connected to the Colibrì.
PRIMING PHASE
Perform the priming with an isotonic solution NaCl 0.9%.
Priming shall be performed first by following the procedure indicated in the ECMOLIFE Centrifugal Pump
instructions for use, ensuring all components are correctly deareated.
5.3.1. AUTOMATIC PRIMING AND SELF LEARNING PROCEDURE – ADULT/PAEDIATRIC CONFIGURATION
After manual priming is possible to continue with the automatic priming operation which also perform a centrifugal
pump self-learning procedure in order to improve the centrifugal pump safety and performances.
Warning
Automatic Priming must be executed once the circuit has been correctly filled with priming
solution.
Empty circuit might cause disposable centrifugal pump rupture or device malfunctions.
In case of automatic priming failure, Eurosets Disposable Centrifugal Pump, Tubing Sets and
Oxygenators instruction for use shall be consulted to perform manual priming.

START / STOP PROCEDURE

You can select "START PRIMING" in order to start the
automatic procedure.
ECMOLIFE will set and hold for few minutes different
RPM targets within the range 0-5000RPM.
During this phase, the circuit will be automatically
primed and the centrifugal pump impeller will be
further verified in a self-learning procedure.
In case of emergency, procedure can be stopped and
restarted many time as necessary.
If the procedure is interrupted and restarted while in
progress, already verified set-point won't be executed
again.
Priming procedure requires a maximum of 4 minutes.
Once completed, it is possible to continue with the
patient connection.