Download Print this page

Alarms Priority; Alarm Pause; Alarm Acknownledge; Alarms Table - Eurosets COLIBRI SYSTEM Instructions For Use Manual

Advertisement

6.3. ALARMS PRIORITY

Alarms activation depends strictly from the priority ranking of parameter involved, according to the following table.
Priority Ranking
1
2
3
4
5
6

6.4. ALARM PAUSE

When an alarm is activated, on the main buttons section, the button is displayed in the
"variable" button which allows the alarm sound to be silenced for 2 minutes.
The visual alarm status is kept active until it is resolved or the alarm is deactivated for that
specific parameter.

6.5. ALARM ACKNOWNLEDGE

When an alarm is activated, on the main buttons section, the button is displayed in the
"variable" button which allows the alarm sound to be silenced permanently.
The visual alarm status is kept active until it is resolved or the alarm is deactivated for that
specific parameter.
Warning
When power is lost for less than or equal to 30 s, the ALARM SETTINGS prior to the power loss is
restored automatically.

6.6. ALARMS TABLE

Refer to ANNEX 1 – PRIMARY: Alarms table for the complete list of alarm messages.
en
Parameter
Blood Flow
Pump Speed
Bubble
Pdrain
Pin
Pout
Priority level
High
High
High
Medium
Medium
Medium
Color
Red
Red
Red
Yellow
Yellow
Yellow
31

Advertisement

loading
Need help?

Need help?

Do you have a question about the COLIBRI SYSTEM and is the answer not in the manual?