15. ANNEXES
15.1.
ANNEX 1: ALARMS TABLE
When connected to Ecmolife, alarms will appear and be handled by the Ecmolife screen. Only the alarm string will be displayed on Colibrì, and the top bar will color according
to the priority of the alarm itself.
On Ecmolife, Colibrì technical (T) alarms will be identified by a "C" within the alarm list, and by the string "COLIBRI ALARM" on the main alarm screen
Physiological
(P) Technical
Alarm displayed
(T)
Internal battery
P
not charging
Internal Battery
P
in reserve
Internal Battery
P
depleted
Battery one in
P
reserve
Internal one
P
depleted
Battery two in
P
reserve
Battery two
P
depleted
P
Low Blood Flow
P
High Blood Flow
P
Low RPM
en
Alarm description
The device is not connected to main
power supply
Internal battery level lower than 70%
Internal battery level lower than 20%
External battery one level lower than 70%
External battery one level lower than 20%
External battery two level lower than 70%
External battery two level lower than 20%
Measured blood flow is less than
minimum value sets in Alarm Page
Measured blood flow is bigger than
maximum value sets in Alarm Page
Measured/selected RPM value is less
than minimum value sets in Alarm Page
Priority
Corrective action
Verify the power module is correctly connected to the device;
High
otherwise if you what to proceed without main power connect
press the acknowledge button and continue the treatment.
Press the acknowledge button and continue the treatment. If you
Medium
desire connect the device to power module or to external battery.
Press the acknowledge button and continue the treatment.
Medium
Connect, in a few time, the device to power module or to external
battery.
Press the acknowledge button and continue the treatment. If you
Medium
desire connect the device to power module or to external battery
two.
Press the acknowledge button and continue the treatment. If you
Medium
desire connect the device to power module or to external battery
two.
Press the acknowledge button and continue the treatment. If you
Medium
desire connect the device to power module or to external battery
one.
Press the acknowledge button and continue the treatment. If you
Medium
desire connect the device to power module or to external battery
one.
Check no occlusions or any other case which may compromise the
flow are presents.
High
Ensure appropriate flow alarms are set.
Do not increase RPM without confirming adequate blood volume is
available.
Check any case which may compromise the flow are presents.
High
Ensure appropriate flow alarms are set.
Consider to decreased RPM to decrease the flow if appropriate.
Check any case which may compromise the flow are presents.
Ensure appropriate RPM alarms are set.
High
Consider to increase LPM (if LPM control is enable) to increase
RPM if appropriate.
53