P
RPM Limitation
T
Speed Error
T
Overcurrent Error
Pumping Body
T
not inserted
T
Flow sensor error
Flow sensor
P
disconnected
Flow Sensor
T
Disconnected
en
Calibration is running
The setting speed is different from
measured speed
An overcurrent condition is detected
Body pump is detected as not inserted
Blood Flow management error
Blood Flow probe is detected as not
inserted
An error is detected on Blood Flow board
Press the alarm acknowledge button and wait the end of the
calibration procedure.
Medium
At the end of calibration procedure the yellow bar disappear and it
is possible research higher RPM/LPM.
An audible alarm will sound and the System will continue to
operate. Press the alarm acknowledge button, if the visual alarm
message does not disappear, clamp the return tubing, stop the
High
Pump and switch to the backup COLIBRI', Motor and Flowmeter.
Resume support. Record the alarm message and contact Technical
Assistance.
Press the alarm acknowledge button and check that an automatic
restarting procedure is executed by the device. The automatic
restarting procedure restore the previously pump speed reduced by
10%. If this procedure doesn't start, check the correct functioning
of the Centrifugal Disposable Pump and set new pump speed
value. If no anomalies are perceived, continue the treatment and
contact Technical Assistance. If the anomaly persist, clamp the
return tubing and switch to the backup COLIBRI', Motor and
Flowmeter and contact Technical Assistance.
Press the alarm acknowledge button and check that an automatic
restarting procedure is executed by the device. The automatic
restarting procedure restore the previously pump speed reduced by
10%. If this procedure doesn't start, verify that the Centrifugal
High
Pump is fully inserted into motor driver and set a new pump speed
value. If the anomaly persist, clamp the return tubing and switch to
the backup COLIBRI', Motor and Flowmeter. Resume support.
Record the alarm message and contact Technical Assistance.
Check the Flow Probe connection to the tubing set.
This error may occur after the LPM set-point modification in case of
the device can't reach the desired flow within two seconds.
In this case be sure no occlusions or any other case which may
High
compromise the flow are presents.
In case of persistent error, disconnect and reconnect the flowmeter
to its relevant connector. If the problem persists, connect the
backup flowmeter.
Check the Flow Probe connection on front of Console.
If necessary, try to disconnect and reconnect the flowmeter to its
High
relevant connector. Press the alarm acknowledge button. Switch to
the backup Flow Probe if the alert message repeats.
Check the Flow Probe connection on front of Console. In case of
persistent error, try to disconnect and reconnect the flowmeter to
its relevant connector and repeat the diagnostic phase. If the
High
anomaly persist, clamp the return tubing and switch to the backup
COLIBRI', Motor and Flowmeter. Resume support. Record the alarm
message and contact Technical Assistance.