Electrical specifications
Power supply
Mains insulation
Power supply cord
Supply voltage
Absorbed input power
External Accessible Fuses
(Appliance Inlet)
Applied part (IEC 60601-1)
Classification of degree of
protection against electrical
hazards (IEC 60601-1)
Degree of protection against
penetration of liquids
48
•
Connection to electric power distribution system through power cord Colibrì
Power Module EU6716
•
Electromagnetic environment: hospital - dedicated supply systems
•
Connection to continuous current power source (24 Vdc)
•
Internal Battery: Li-Ion rechargeable batteries, 25.9 V, 10.4 Ah
•
Internal Battery: NiMH rechargeable battery, 7.2 V, 4500 mAh
•
External Battery: AccuPower, model: Lithium Battery 7S4P
•
Bipolar emergency switch
•
Power supply cord detachable with lock system and mains plug
•
Default cable: Shurter 3-101-795
•
Mains plug: CEE 7/7 (E/F) - straight
•
Rated Voltage: 250 V~
•
Length: 3 m
•
Colibrì Power Module: 100-240 Vac / 0.9-1.8 A (Test condition: 230/115 V)
/ 210 VA 50 – 60 Hz
•
Integrated battery or DC External Power Supply: 24-29.4Vdc / 5A / 147W
190 VA
T4 A – 250 V – 5x20 mm
Tubing Set including ECMOLIFE Centrifugal Pump
Class II type CF
IP33